Find Infant Incubator, Baby Incubator, Incubator For Babies on Industry Directory, Reliable Manufacturer/Supplier/Factory from China.

Inquiry Basket (0)
Home > Products > Gynaecology/Neonatal Equipment > Infant Incubator > High Grade Medical Equipment Infant Incubator

High Grade Medical Equipment Infant Incubator

Basic Info

Model No.:  BI-1000

Product Description

Model NO.: BI-1000 Servo Control Mode: Air Temp Audible & Visual Alarms: Power Failure, Sensor Failure, Higher & Lower Temp Transport Package: Standard Packaging for Export Color: White Instant Parameter Displaying: Set Temp, Air Temp and Heat Power Rate Trademark: perlong Origin: Jiangsu, China High Grade Medical Equipment Infant Incubator
Specifications
Model BI-3000 BI-1000 BI-800
Canopy Specifications
Mattress (canopy) 450mm 450mm 450mm
Access Port 6 (2 Iris ports) 6 (2 Iris ports) 4 (2 Iris ports)
Self-sealing Tube Ports 4 4 2
Front Access Panel 826x310mm 826x310mm 826x310mm
Mattress Tray Size 630x360mm 630x360mm 630x360mm
Technical Specifications
Temperature Mode (Servo Controlled) air, infant air, infant air
Air Temperature Adjustment Range 20°C~37°C, 37~39°C 20°C~37°C, 37~39°C 20°C~37°C
Infant Temperature Adjustment Range 30°C~37°C, 37~38°C 30°C~37°C, 37~38°C N/A
Infant Sensor Accuracy ≤0.3°C ≤0.3°C N/A
Relative Humidity Display Range 5%~99% N/A N/A
Temperature Fluctuation 0.5°C 0.5°C 0.5°C
Uniformity of Mattress Temperature 0.8°C 0.8°C 0.8°C
Time to Warm Up 30 min 30 min 30 min
Noise Level within Canopy 55dB 55dB 55dB
Alarms
Power Failure Alarm Yes Yes Yes
Fan Failure Alarm Yes Yes Yes
Sensor Failure Alarm Yes Yes Yes
Air Temperature Deviation Alarm ±3.0°C ±3.0°C ±3.0°C
Infant Temperature Deviation Alarm ±1.0°C ±1.0°C N/A
Air Temperature Exceeds set Limit Alarm 38.0°C; 40.0°C 38.0°C; 40.0°C 38.0°C
Infant Temperature Exceeds set Limit alarm 38.0°C; 39.0°C 38.0°C; 39.0°C N/A
System Failure Alarm Yes Yes N/A
Mute Alarm Function Yes Yes N/A
Operational Environment
Temperature within Operational Environment +20°C~+30°C +20°C~+30°C +20°C~+30°C
Relative Humidity within Operational Environment 5%~99% 5%~99% 5%~99%
Velocity of Air Flow within Operational Environment <0.3m/s <0.3m/s <0.3m/s
Other Specifications
RS-232 connector Yes Yes N/A
Tilt of Infant Bed (Adjustment Range) ±10° step-less ±10° step-less N/A
Maximum Load of Mattress Tray 5kg 5kg 5kg
Maximum Load of IV Stand 2kg 2kg 2kg
Maximum Load of Cabinet 6kg 6kg N/A

Contact us if you need more details on Incubator. We are ready to answer your questions on packaging, logistics, certification or any other aspects about Infant IncubatorMedical Equipment. If these products fail to match your need, please contact us and we would like to provide relevant information.

Product Categories : Gynaecology/Neonatal Equipment > Infant Incubator